title: Can CryoPen be sterilised?
description: CryoPen does not normally require sterilisation when used correctly, as it should not contact the client. If the applicator accidentally touches the client, it can be autoclaved at 134°C for 4 minutes or 121°C for 20 minutes.
---

[Skip to content](https://ivanmedportal.com/knowledge-base/what-question-is-your-artcan-cryopen-be-sterilisedicle-answering#main-content)

English

Show submenu for translations

[More support](https://ivanmedportal.com/knowledge-base/kb-tickets/new?hsLang=en) [Customer portal](https://ivanmedportal.com/customer?hsLang=en)

![Ivan-Med-Logo](https://ivanmedportal.com/hs-fs/hubfs/Ivan-Med-Logo.png?width=200&height=63&name=Ivan-Med-Logo.png)

- [Knowledge Base](https://ivanmedportal.com/knowledge-base)
- [Back To Site](https://ivanmed.com/)
- [Sign Out](https://ivanmedportal.com/_hcms/mem/logout)

Open main navigation

Close main navigation

- [Knowledge Base](https://ivanmedportal.com/knowledge-base)
- [Back To Site](https://ivanmed.com/)
- [Sign Out](https://ivanmedportal.com/_hcms/mem/logout)
- English
  
  Show submenu for translations
- [More support](https://ivanmedportal.com/knowledge-base/kb-tickets/new)
- [Customer portal](https://ivanmedportal.com/customer)

 How can we help you?

- There are no suggestions because the search field is empty.

1. [Help Center](https://ivanmedportal.com/knowledge-base?hsLang=en)
2. [Cryopen](https://ivanmedportal.com/knowledge-base/cryopen?hsLang=en)
3. [Technical Troubleshooting](https://ivanmedportal.com/knowledge-base/cryopen?hsLang=en#technical-troubleshooting)

# Can CryoPen be sterilised?

## The content explains that CryoPen normally does not require sterilisation when used correctly because the device should not contact the client.

[CryoPen Maintenance](https://ivanmedportal.com/hubfs/CRYOPENmaintenance_09_2024.pdf?hsLang=en)

CryoPen does not normally require sterilisation because, when used correctly, the device does not come into contact with the client’s skin or lesion.

If the **applicator accidentally comes into contact with the client**, it can be cleaned, disinfected and autoclaved in accordance with the manufacturer’s instructions:

- **134°C for 4 minutes**, or
- **121°C for 20 minutes**.

The main CryoPen instrument must **not be immersed in disinfecting liquids**, as this may damage the internal switches. External surfaces can be wiped with a non-corrosive disinfecting solution. Please note, never clean the thread of frontend and backend. The existing coating should be protected. If the lubricant is wiped off, that creates friction which affects the piercing of the cartridge. 

Always follow the manufacturer’s instructions and the disinfectant manufacturer’s recommended concentration and contact time.

- [imedpen & microneedling](https://ivanmedportal.com/knowledge-base/imedpen-microneedling?hsLang=en#main-content)
  
  
  
  
  
    - [Treatment questions](https://ivanmedportal.com/knowledge-base/imedpen-microneedling?hsLang=en#treatment-questions)
    - [Technical questions](https://ivanmedportal.com/knowledge-base/imedpen-microneedling?hsLang=en#technical-questions)
- [Cryopen](https://ivanmedportal.com/knowledge-base/cryopen?hsLang=en#main-content)
  
  
  
  
  
    - [Treatment questions](https://ivanmedportal.com/knowledge-base/cryopen?hsLang=en#treatment-questions)
    - [Technical Troubleshooting](https://ivanmedportal.com/knowledge-base/cryopen?hsLang=en#technical-troubleshooting)
- [Training Information](https://ivanmedportal.com/knowledge-base/training-information?hsLang=en)
- [Order](https://ivanmedportal.com/knowledge-base/order?hsLang=en)
- [Loyalty Scheme](https://ivanmedportal.com/knowledge-base/loyalty-scheme?hsLang=en)
- [Terms and Conditions](https://ivanmedportal.com/knowledge-base/terms-and-conditions?hsLang=en)
- [Juliette Armand](https://ivanmedportal.com/knowledge-base/juliette-armand?hsLang=en)

- [Knowledge Base](https://ivanmedportal.com/knowledge-base)
- [Back To Site](https://ivanmed.com/)
- [Sign Out](https://ivanmedportal.com/_hcms/mem/logout)

## Ivan med

Copyright © 2025, Ivan med